292
Comment:
|
1509
|
Deletions are marked like this. | Additions are marked like this. |
Line 3: | Line 3: |
For more information, type in MATLAB {{{#!highlight matlab numbers=off help ScannerSynchClass }}} {{{#!wiki caution N.B.: National Instruments PCI 6503 card does not monitor pulses in the background, so you have to make sure that you wait for any pulse before it comes! }}} |
|
Line 7: | Line 16: |
== Obtain ScannerSynch == * ScannerSynch is available on all relevant PCs (StimPCs, "mimic" PC) * From your desktop PC (Windows), you can also have a copy from `\\cbsu\data\Group\Imaging-Local\cbu\matlab` * You can also download it from [[attachment:ScannerSynchClass.m|here]] (right-click and "Save As...") == Using ScannerSynch == === Initialise === * Default. Use only if device (scanner, pulse emulator) is connected (only in StimPCs and mimic PC)! . {{{#!highlight matlab numbers=off SSO = ScannerSynchClass; }}} . {{{#!highlight matlab numbers=off SSO = ScannerSynchClass(1); % emulate scanner synch pulse }}} . {{{#!highlight matlab numbers=off SSO = ScannerSynchClass(0,1); % emulate button box }}} . {{{#!highlight matlab numbers=off SSO = ScannerSynchClass(1,1); % emulate scanner synch pulse and button box }}} === Close === . {{{#!highlight matlab numbers=off SSO.delete; }}} |
ScannerSynch
ScannerSynch provides an interface for National Instruments PCI 6503 card used to capture all pulses (scanner synch pulse and button pulses alike).
For more information, type in MATLAB
help ScannerSynchClass
N.B.: National Instruments PCI 6503 card does not monitor pulses in the background, so you have to make sure that you wait for any pulse before it comes!
Any problems, report them to Tibor Auer.
Obtain ScannerSynch
ScannerSynch is available on all relevant PCs (StimPCs, "mimic" PC)
From your desktop PC (Windows), you can also have a copy from \\cbsu\data\Group\Imaging-Local\cbu\matlab
You can also download it from here (right-click and "Save As...")
Using ScannerSynch
Initialise
- Default. Use only if device (scanner, pulse emulator) is connected (only in StimPCs and mimic PC)!
SSO = ScannerSynchClass;
SSO = ScannerSynchClass(1); % emulate scanner synch pulse
SSO = ScannerSynchClass(0,1); % emulate button box
SSO = ScannerSynchClass(1,1); % emulate scanner synch pulse and button box
Close
SSO.delete;