Diff for "VectorviewDescription" - Meg Wiki
location: Diff for "VectorviewDescription"
Differences between revisions 5 and 6
Revision 5 as of 2010-04-15 11:31:55
Size: 981
Editor: YaaraErez
Comment:
Revision 6 as of 2010-04-15 11:36:37
Size: 874
Editor: YaaraErez
Comment:
Deletions are marked like this. Additions are marked like this.
Line 12: Line 12:
We routinely use a 70-electrode EEG set-up. This is the electrode layout:
Line 13: Line 14:
== FreewarePackages ==

'''[http://imaging.mrc-cbu.cam.ac.uk/meg/SpmAnalysis SPM]''', developed at the [http://www.fil.ion.ucl.ac.uk/spm/ FIL] and at the [http://imaging.mrc-cbu.cam.ac.uk/meg/SpmAnalysis CBU].
attachment:ElectrodeLayout.jpg

The Elekta Neuromag Vectorview System

Technical description

Our Vectorview systems contains 102 magnetometers and 204 planar gradiometers. Magnetometers consist of a single coil which measures the magnetic flux perpendicular to its surface. Planar gradiometers consist of a "figure-of-eight"-type coil configuration. The measured signal is the difference between the two loops of the "eight", or the spatial gradient. There are two orthogonal planar gradiometers at each location. This arrangement is illustrated in the following figure (black: magnetometer; grey and white: two gradiometers).

attachment:MagsGrads.jpg

This is the whole sensor layout, including the sensor labels used by the Neuromag software:

attachment:SensorLayout.jpg

We routinely use a 70-electrode EEG set-up. This is the electrode layout:

attachment:ElectrodeLayout.jpg

CbuMeg: VectorviewDescription (last edited 2024-04-18 13:58:10 by OlafHauk)