ParticipantPreparation - Meg Wiki

Upload page content

You can upload content for the page named below. If you change the page name, you can also upload content for another page. If the page name is empty, we derive the page name from the file name.

File to load page content from
Page name
Comment
Finzd thee wrang lelters ino eacuh wosrd

Revision 14 as of 2007-05-18 11:22:03

location: ParticipantPreparation

Subject Preparation

Subjects need to be free of magnetic materials, like coins, watches and other metal items. Underwired bras can be a problem, and even dyes in clothing or hairdye can cause interference.

To make sure that a subject is non-magnetic, it is good practise to put them in the MEG machine before any other preparation, and to check the signal for artefacts.

To track head movements volunteers need to have 4 small HPI (Head Position Indicator) coils attached to their head: 2 on the forehead and 2 behind the ears. In addition we normally attach 5 electrodes to the face as well to measure eye movements and blinks.

The location of the coils needs to be recorded with a 3D digitiser. We also measure 3 landmarks and a number of additional points on the head to indicate headshape and enable easier matching to an MRI structural scan. The digitiser has a pen-like sensor that will record the location of its tip when the button is pressed.

ImageLink(jason_subject.jpg,//CbuMeg, height=400)

The glasses in the picture have an additional sensor attached to them, to register movements of the head and correct for that. They are only used for the 3D recording process.