RED stands for 'Remote Eye-tracking Device', and this is a remote system. For that reason the precision is limited, but this is a very easy system to use and very comfortable for the subject. The system is also relatively portable and consists of a monitor with added camera unit, a power supply box and a laptop. RED stands for 'Remote Eye-tracking Device', and this is a 50 Hz remote system. For that reason the precision is limited, but this is a very easy system to use and very comfortable for the subject. The system is also relatively portable and consists of a monitor with added camera unit, a power supply box and a laptop.

Eye tracking

The CBU currently has 4 eye trackers. One is in the MEG lab, one in the MRI (from May 2009), and we have two separate eye trackers for use in other locations. All eye trackers are made by SMI and use the same software for controlling the eye tracking hardware, for stimulus presentation and for analysis of the eye tracking data. In addition, all trackers can also be used with E-Prime, if needed.

All 4 eye trackers use 'dark pupil' technology, where the gaze of the eye is tracked by identifying the pupil and the reflection of an infra-red light source on the cornea.

Eye trackers come in two variaties: 'contact' and 'remote'. With contact eye trackers the camera and light source are fixed to the head, or the head is on a chin rest to limit movements relative to the camera. In remote systems the camera and light source are in a fixed location, at some distance from the subject, enabling some head movements. Contact systems are more precise and reliable, in general.

Another important feature of an eye tracker is the frequency. To be able to follow the eye during saccades the minimum frequency is about 200 Hz.

All eye trackers are able to output the x and y coordinates of the screen location the subject is looking at, and the diameter of the pupil. That last value will be 0 during a blink.

MEG eye tracker

This is a 50 Hz system that should theoretically be called a remote eye tracker, but actually behaves more like a contact system as head movement is extremely limited in the MEG helmet. For that reason the camera can actually be zoomed in quite closely and the accuracy of the MEG eye tracker should be reasonably high.

The eye tracker sits on a tripod under the screen, about 1.2 meter from the subject. It is controlled from a separate PC in the control room.

The eye tracking data will be recorded with the MEG/EEG data in the .fif file produced by the MEG acquisition software. To enable this you will have to select the three first MISC channels in your setup.

In addition, the eye tracking data will also be saved on the PC controlling the tracker. This PC is connected to the Elekta trigger box, so will also record the triggers coming from the stimulus presentation machine and all responses from the subject.

MRI eye tracker

The MRI eye tracker will arrive in May. It will be a system comparable to that in the MEG.

RED eye tracker

RED stands for 'Remote Eye-tracking Device', and this is a 50 Hz remote system. For that reason the precision is limited, but this is a very easy system to use and very comfortable for the subject. The system is also relatively portable and consists of a monitor with added camera unit, a power supply box and a laptop.

Hi-speed eye tracker

This is a contact system that will allow frequencies of up to 1250 Hz for monocular eye tracking and 500 Hz for binocular tracking. The Hi-speed system is extremely precise and reliable, but requires the subject to put their head on a chin rest.

Eye tracking in general

Eye tracking is not as easy as it might look, and there's also quite a bit of variability between subjects. Most problems can be solved and with the vast majority of people you should be able to acquire decent quality eye tracking data.

Calibration

The first thing that needs to be done is a calibration. This is an automatic procedure that will adjust the main parameters to the specific subject. Calibration shouldn't take more that 20-30 seconds. If the subject leaves the setup, during a break, you will have to calibrate again at the beginning of the next block.

Problems

Mascara can make eye tracking impossible, as the software will interpret the black regions in the picture as the pupil, and mascare is very black too. Mascare will alwasy have to be removed, and it is best to ask people not to wear any mascara when they participate in an experiment involving eye tracking.

Drooping eye lids can be a problem, as they can partly obscure the pupil. This is more common with older people. The solution is to move the camera to a lower position, so that the eye is filmed from below.

Left-right difference. Eye tracking data will always be less reliable and distorted towards the edges of the screen, and the problems are more serious for the eye on the opposite side. Try to present your stimuli is the center section of the screen, if possible.

Monocular versus binocular

Some of our eye trackers offer the choice between monocular and binocular eye tracking. There is no real consensus about which one is better. The RED system will always record binocular data, and the MEG and MRI systems will usually record monocular data. The hi-speed system can be switched between the two modes very quickly.

When using an eye tracker in monocular mode, the next question is which eye to use. There are two possibilities: using the same eye for all subjects, or using the dominant eye for each subject. the dominant eye can be found by asking the subject to look through a small hole in a card. People do this with their dominant eye.

The advantage of using the same eye for all subjects is that your data will be more consistent. This is because of the left-right distortion mentioned in the previous section.

The advantage of using the dominant eye is that the data will be less noisy. The non-dominant eye can have the tendency to make 'glissades', small eye movements to align itself with the dominant eye at the end of a saccade. These will not be present to the same extent in the dominant eye.