Stuff Three different TR are available for the Quiet EPI sequence, TR=4s (TE=44; 32 slices; 24db quieter than standard); TR=2.7s (TE=44; 32 slices; 21db quieter); TR=2.2s (TE=38ms; 32 slices; 8db quieter).[is these right? the sequences differ wrt noise level as compared to normal]

Re dropout and distortion: They scale up with TR, so would be best to use TR=2.2s to avoid these artifacts. However, this does not seem to be a good sequence for regions that are susceptible to the dropout artifact, e.g. inferior temporal and frontal lobes. In addition, Matt pointed out there seems to be some problem with imaging the cerebellum. Reason for this is unclear.

Re distortion: Field-maps could be used for undistortion. Action point: Rhodri would liaison with Gayaneh, Matt, and Jonathan.
Re wrap around: After these images were acquired Marta received a new sequence from the developers, which seems to avoid the wrap around effect. Which sequence is this? It this the one that we’ll use from now on?

Re ghosting: If could be helped by adopting a small field of view. Or, one could use a normal field of view and then drop out the data around the edges. Also, careful orientation of the slices might avoid the problem of ghosting. However whether this is feasible depends on which region is of interest for the data acquisition.

Re inter-subject variability: The shimming of the scanner could have to do with the inter-subject variability.
General re data quality: There will be piloting of the sequence. A number of acquisitions are going to be acquired from subjects that are being run at CBU from week 04/10/2010. Two sessions of 20s each. Field-maps would be acquired for each participant Action point: Marta would organize these acquisitions and feed back to the group. Some parameters remain to be decided?

The data would be pre-processed with AA to explore the data quality with regards to the artifacts. Undistortion could be applied to see whether this improves the data quality. The mean EPI image and tsdiffana can be used to explore data quality. Lower slices are expected to have higher slice variability. The piloting could help to see whether the inter-subject variability is affected by the shimming of the scanner, by looking at whether noise/artifacts change over the acquisition time across subjects.

Re BOLD sensitivity: It is not clear whether and/or how these problems affect the sequence’s BOLD sensitivity. But, see paper by Jonathan Peele et al 2010. Action point: Lorina and Annika can pilot the quiet EPI sequence further with a functional task, but this depends on the slot availability. There are none to be used for method slots this week. Would it be possible to use physics morning slots for this? If so, would the shimming be especially a problem for these slots? Our first pilot, acquired around 9am had more artifacts than the subsequent one.

Describe UsingTheSilentSequence here.

Main title

Several issues were encountered with using the Quiet EPI.

Artifacts

* Dropout

* Distortion (Some squashing of the brain. Similar to old Brooker data)

* Wrap around (in some cases the nose ended up at the back of the brain)

* Ghosting (in the form of streaks of image at the back of the head. Also there is what looks like tiger stripes on the brain surface, around the edges.)

* High inter-subject variability in the data quality (in two different datasets, the subject acquired earlier in time had worse data than the later subject)

Three different TR are available for the Quiet EPI sequence, TR=4s (TE=44; 32 slices; 24db quieter than standard); TR=2.7s (TE=44; 32 slices; 21db quieter); TR=2.2s (TE=38ms; 32 slices; 8db quieter).[is these right? the sequences differ wrt noise level as compared to normal]

Re dropout and distortion: They scale up with TR, so would be best to use TR=2.2s to avoid these artifacts. However, this does not seem to be a good sequence for regions that are susceptible to the dropout artifact, e.g. inferior temporal and frontal lobes. In addition, Matt pointed out there seems to be some problem with imaging the cerebellum. Reason for this is unclear.

Re distortion: Field-maps could be used for undistortion. Action point: Rhodri would liaison with Gayaneh, Matt, and Jonathan. Re wrap around: After these images were acquired Marta received a new sequence from the developers, which seems to avoid the wrap around effect. Which sequence is this? It this the one that we’ll use from now on?

Re ghosting: If could be helped by adopting a small field of view. Or, one could use a normal field of view and then drop out the data around the edges. Also, careful orientation of the slices might avoid the problem of ghosting. However whether this is feasible depends on which region is of interest for the data acquisition.

Re inter-subject variability: The shimming of the scanner could have to do with the inter-subject variability. General re data quality: There will be piloting of the sequence. A number of acquisitions are going to be acquired from subjects that are being run at CBU from week 04/10/2010. Two sessions of 20s each. Field-maps would be acquired for each participant Action point: Marta would organize these acquisitions and feed back to the group. Some parameters remain to be decided?

The data would be pre-processed with AA to explore the data quality with regards to the artifacts. Undistortion could be applied to see whether this improves the data quality. The mean EPI image and tsdiffana can be used to explore data quality. Lower slices are expected to have higher slice variability. The piloting could help to see whether the inter-subject variability is affected by the shimming of the scanner, by looking at whether noise/artifacts change over the acquisition time across subjects.

Re BOLD sensitivity: It is not clear whether and/or how these problems affect the sequence’s BOLD sensitivity. But, see paper by Jonathan Peele et al 2010. Action point: Lorina and Annika can pilot the quiet EPI sequence further with a functional task, but this depends on the slot availability. There are none to be used for method slots this week. Would it be possible to use physics morning slots for this? If so, would the shimming be especially a problem for these slots? Our first pilot, acquired around 9am had more artifacts than the subsequent one.