Diff for "MrDevelopers" - MRC CBU Imaging Wiki
location: Diff for "MrDevelopers"
Differences between revisions 5 and 14 (spanning 9 versions)
Revision 5 as of 2009-11-11 15:00:04
Size: 2064
Editor: RhodriCusack
Comment:
Revision 14 as of 2009-11-11 15:09:42
Size: 2299
Editor: RhodriCusack
Comment:
Deletions are marked like this. Additions are marked like this.
Line 2: Line 2:
If you are developing IDEA or ICE sequences for the scanner at the CBU, please follow this practice. If you are developing IDEA or ICE sequences for the scanner at the CBU, please follow this practice. 
Line 4: Line 4:
== Virus prevention == === Virus prevention ===
Line 9: Line 9:
== Acoustic resonances == === Scanner settings ===
 * Don't change any settings on the scanner (e.g., using the ideacmdtool) that will disrupt regular scanning in any way.
 * If you're not sure, ask

=== Acoustic resonances ===
Line 12: Line 16:
== Sequence version control ==
Try not to:
 * Overwrite any sequence that someone else is using without letting them know
 * C
hange any sequence that is already in use (even in a way that improves it) without being absolutely sure none of the users will have a problem. I would suggest only allowing users to use "releases" of software, and not deleting earlier releases as new ones come along. A suggested naming convention is EPI_ISSS_r01, EPI_ISSS_r02 and so on
 * Try to keep any users of your system, and the wiki record, updated so that everyone is aware of the strengths/pitfalls/versions of your sequences as you discover them.
=== Sequence version control ===
 * Try not to overwrite any sequence that some other developer is using.
 * Don't change any sequence that is already in use (even in a way that improves it) without being absolutely sure none of the users will have a problem. I would suggest only allowing users to use "releases" of software, and not deleting earlier releases as new ones come along. A suggested naming convention is EPI_ISSS_r01, EPI_ISSS_r02 and so on
 * Keep users of your sequence, and the wiki record, updated so that everyone is aware of the strengths/pitfalls/versions of your sequences as you discover them.
 * Keep the radiographers aware of new sequences and releases
Line 18: Line 22:
== Software upgrades == === Software upgrades ===
Line 20: Line 24:
 * Make upgrades available as soon as is practical.  * Try to make upgrades available when practical.
Line 22: Line 26:
== Working in the facility - local rules ==
 * Don't ever go in the magnet room unless you are accompanied by a radiographer or scanning buddy (who must have done the safety course).
=== Working in the facility - local rules ===
 * Don't ever go in the magnet room unless there is a radiographer or scanning buddy (who must have done the safety course) also in the facility.

Developing New MR Sequences: Good Practice Guide

If you are developing IDEA or ICE sequences for the scanner at the CBU, please follow this practice.

Virus prevention

Viruses have been common on other machines in the scanner room, and we must take precautions to stop these ever being transferred to the scanner.

  • Only use the two USB memory sticks labelled "Scanner only" for transferring sequences to the scanner.
  • Only put these USB sticks into CBU maintained computers that are running Anti-Virus software (NOT the stimulus delivery machines, which do not run antivirus and are often infected).

Scanner settings

  • Don't change any settings on the scanner (e.g., using the ideacmdtool) that will disrupt regular scanning in any way.
  • If you're not sure, ask

Acoustic resonances

Sequences that oscillate the gradients at particular frequencies can set up resonances that can damage the system. Please check all new sequences before using them on the Tim Trio. (add instructions for doing this here)

Sequence version control

  • Try not to overwrite any sequence that some other developer is using.
  • Don't change any sequence that is already in use (even in a way that improves it) without being absolutely sure none of the users will have a problem. I would suggest only allowing users to use "releases" of software, and not deleting earlier releases as new ones come along. A suggested naming convention is EPI_ISSS_r01, EPI_ISSS_r02 and so on
  • Keep users of your sequence, and the wiki record, updated so that everyone is aware of the strengths/pitfalls/versions of your sequences as you discover them.
  • Keep the radiographers aware of new sequences and releases

Software upgrades

  • In advance of any scanner software upgrade, work out whether it will break your sequence (probably!) and then alert your users that the sequence will stop working on this date.
  • Try to make upgrades available when practical.

Working in the facility - local rules

  • Don't ever go in the magnet room unless there is a radiographer or scanning buddy (who must have done the safety course) also in the facility.
  • Don't ever put a human volunteer in the scanner without a radiographer present
  • Read the SOPs