Diff for "ImagingSequences" - CBU MRI facility Wiki
location: Diff for "ImagingSequences"
Differences between revisions 1 and 9 (spanning 8 versions)
Revision 1 as of 2006-08-01 14:19:48
Size: 1862
Editor: devel03
Comment:
Revision 9 as of 2007-02-15 14:57:44
Size: 2164
Comment:
Deletions are marked like this. Additions are marked like this.
Line 6: Line 6:
|| Acquisition time || 2 s (per volume) ||  || Acquisition time || 2 s (per volume) ||
Line 11: Line 11:
|| Slice order || Interleaved (1, 3, 5... 2, 4, 6) || || Slice order || Descending (32,31,...,1) ||
Line 18: Line 18:
|| Bandwidth || 2232 Hz/Px || || Bandwidth || 2232 Hz/Px ||
||
Echo spacing || 0.51 ms ||
Line 23: Line 24:

=== Slice order ===

The slice order is interleaved: 1, 3, 5... 2, 4, 6. To put this into SPM directly when asked for the slice order write [1:2:31 2:2:32]. This is the default for aa_ver1
Line 40: Line 37:
|| Parallel imaging || GRAPPA ||
|| Acceleration factor PE: 2 ||
|| Reference lines PE: 24 ||
|| Parallel imaging || GRAPPA [[BR]] Acceleration factor PE: 2 [[BR]] Reference lines PE: 24 ||

== CBU_TSE ==

|| Description || T2-weighted structural MRI ||
|| Acquisition time || 1 min 38 s ||
|| Slices || 29 ||
|| Slice gap || 30% ||
|| Slice thickness || 4 mm ||
|| Type || 2D ||
|| FOV || 220 mm x 220 mm ||
|| Matrix || 320 x 320 ||
|| Resolution || 0.7 x 0.7 mm ||
|| TR || 5060 ms ||
|| TE || 102 ms ||
|| Flip angle (refoc. pulse) || 140 deg ||
|| Parallel imaging || GRAPPA [[BR]] Acceleration factor PE: 2 [[BR]] Reference lines PE: 26 ||

Standard imaging sequences

CBU_EPI

Description

Standard EPI sequence for fMRI

Acquisition time

2 s (per volume)

Type

2D

Slices

32

Slice gap

25%

Slice thickness

3 mm

Slice order

Descending (32,31,...,1)

FOV

192 mm x 192 mm

Matrix

64 x 64

Resolution

3 mm x 3mm

TR

2000 ms

TE

30 ms

Flip angle

78 deg

Bandwidth

2232 Hz/Px

Echo spacing

0.51 ms

Trigger pulse timing

To within a few ms, the stimulus delivery trigger pulse happens 35 ms before the acquisition of the first scan (centre of k space) of each volume. Note there is a latency of 20 ms in the visual display. The trigger pulse length is 1 ms.

CBU_MPRAGE

Description

High-resolution structural MRI

Acquisition time

4 min 16 s

Type

3D

FOV

256 mm x 240 mm x 160 mm

Matrix

256 x 240 x 160

Resolution

1 mm isotropic

TR

2250 ms

TI

900 ms

TE

2.99 ms

Flip angle

9 deg

Parallel imaging

GRAPPA BR Acceleration factor PE: 2 BR Reference lines PE: 24

CBU_TSE

Description

T2-weighted structural MRI

Acquisition time

1 min 38 s

Slices

29

Slice gap

30%

Slice thickness

4 mm

Type

2D

FOV

220 mm x 220 mm

Matrix

320 x 320

Resolution

0.7 x 0.7 mm

TR

5060 ms

TE

102 ms

Flip angle (refoc. pulse)

140 deg

Parallel imaging

GRAPPA BR Acceleration factor PE: 2 BR Reference lines PE: 26

CBU_Fieldmapping

Description

Magnetic field mapping sequence. Required for BR - undistortion of fMRI data BR - functional sensitivity mapping

Acquisition time

54 s

Type

2D

Slices

32

Slice gap

25%

Slice thickness

3 mm

FOV

192 mm x 192 mm

Matrix

64 x 64

Resolution

3 mm x 3 mm

TR

400 ms

TE

5.19 ms / 7.65 ms

Flip angle

60 deg

None: ImagingSequences (last edited 2013-03-08 10:21:22 by localhost)